FDA Enforcement Class II Ongoing

Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4702

Recall: Z-0469-2025 · Reported November 27, 2024

Enforcement

Recall Number
Z-0469-2025
Event ID
95577
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Hobbs Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 27, 2024
Initiation Date
October 4, 2024
Classification Date
November 15, 2024
Address
8 Spring St, N/A, Stafford Springs, CT, 06076-1505, United States

Description

Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4702

Reason

Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

Code Info

Catalog Number: 4702; UDI-DI: M84947020; Lot Numbers: H08-23-263, H01-24-114.

Distribution

Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.

Quantity

45 units