FDA Enforcement Class II Terminated

Stryker Orthopaedics FEMORAL OFFSET BUSHING TRITHION REVISION INSTRUMENTS REF 6543-1-600;NON-STERILE Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA A Subsidiary of Stryker Corp. Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 Meyzieu Cedex France The Offset Bushings are parts of the Triathlon TS system. They are used during trialing to assess the appropriate magnitude and orientation of offset between the Implant and the intramedullary Stem. The feedback from the Offset Bushing allows for the selection of the corresponding Offset Adaptor, which allows for optimal bone coverage, compensating for differences between the Implant and bone canal axes. Made in USA

Recall: Z-0469-2013 · Reported December 12, 2012

Enforcement

Recall Number
Z-0469-2013
Event ID
63442
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Howmedica Osteonics Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 12, 2012
Initiation Date
July 27, 2012
Classification Date
December 4, 2012
Termination Date
September 30, 2015
Address
325 Corporate Drive, N/A, Mahwah, NJ, 07430, United States

Description

Stryker Orthopaedics FEMORAL OFFSET BUSHING TRITHION REVISION INSTRUMENTS REF 6543-1-600;NON-STERILE Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA A Subsidiary of Stryker Corp. Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 Meyzieu Cedex France The Offset Bushings are parts of the Triathlon TS system. They are used during trialing to assess the appropriate magnitude and orientation of offset between the Implant and the intramedullary Stem. The feedback from the Offset Bushing allows for the selection of the corresponding Offset Adaptor, which allows for optimal bone coverage, compensating for differences between the Implant and bone canal axes. Made in USA

Reason

Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not include rotational markings.

Code Info

Class I, 510(k) exempt Catalog Number: 6543-1-600 Lot Number; NE1WR1

Distribution

Worldwide Distribution.

Quantity