FDA Enforcement
Class II
Terminated
Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualitative and Quantitative Factor Deficiency
Recall: Z-0468-2016
·
Reported December 30, 2015
Enforcement
- Recall Number
- Z-0468-2016
- Event ID
- 72752
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 30, 2015
- Initiation Date
- November 19, 2015
- Classification Date
- December 22, 2015
- Termination Date
- August 15, 2017
- Address
- 511 Benedict Ave, N/A, Tarrytown, NY, 10591-5005, United States
Description
Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualitative and Quantitative Factor Deficiency
Reason
Factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves when used for measurement on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, CS-5100 Systems. This could lead, under certain circumstances, to an erroneously increased result with a Factor VIII Chromogenic Assay.
Code Info
Lot Number 44637, exp. 2016-11-30
Distribution
Worldwide Distribution - US Distribution to the state of : Massachusetts and to the countries of : Austria, Czech Republic, France, Germany, Hungary, Italy, Poland, Serbia, Spain and Sweden.
Quantity
Domestic: 52 kits, Foreign: 623 kits