Proteus 235; Version: PTS-8 versions before PTS-8.7.2

Recall: Z-0466-2022 · Reported January 19, 2022

Enforcement

Recall Number
Z-0466-2022
Event ID
89267
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ion Beam Applications S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 19, 2022
Initiation Date
November 29, 2021
Classification Date
January 7, 2022
Termination Date
February 20, 2024
Address
Chemin Du Cyclotron 3, N/A, Louvain-la-neuve, N/A, N/A, Belgium

Description

Proteus 235; Version: PTS-8 versions before PTS-8.7.2

Reason

Proton Therapy System (PTS) software can be started in the clinical environment (clinical user) while some processes are still running in the test environment (tcs user). Initiating the startup of the clinical processes in the environment is not prevented even if some processes of the test environment (tcs) are still running. It may lead to situations where the system is using test processes without notifying the clinical user. If the versions of the processes are different between both environments, this could have an impact on patient treatments.

Code Info

PAT.108 (US), PAT.110 (US), PAT.113 (US), PAT.114 (EU), SAT.119 (US).

Distribution

IN USA: VA, OK, NJ, TN OUS: Czech Republic.

Quantity

5 devices