FDA Enforcement
Class II
Terminated
Proteus 235; Version: PTS-8 versions before PTS-8.7.2
Recall: Z-0466-2022
·
Reported January 19, 2022
Enforcement
- Recall Number
- Z-0466-2022
- Event ID
- 89267
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ion Beam Applications S.A.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 19, 2022
- Initiation Date
- November 29, 2021
- Classification Date
- January 7, 2022
- Termination Date
- February 20, 2024
- Address
- Chemin Du Cyclotron 3, N/A, Louvain-la-neuve, N/A, N/A, Belgium
Description
Proteus 235; Version: PTS-8 versions before PTS-8.7.2
Reason
Proton Therapy System (PTS) software can be started in the clinical environment (clinical user) while some processes are still running in the test environment (tcs user). Initiating the startup of the clinical processes in the environment is not prevented even if some processes of the test environment (tcs) are still running. It may lead to situations where the system is using test processes without notifying the clinical user. If the versions of the processes are different between both environments, this could have an impact on patient treatments.
Code Info
PAT.108 (US), PAT.110 (US), PAT.113 (US), PAT.114 (EU), SAT.119 (US).
Distribution
IN USA: VA, OK, NJ, TN OUS: Czech Republic.
Quantity
5 devices