FDA Enforcement Class II Terminated

NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

Recall: Z-0466-2013 · Reported December 12, 2012

Enforcement

Recall Number
Z-0466-2013
Event ID
63553
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips And Neusoft Medical Systems Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 12, 2012
Initiation Date
October 18, 2012
Classification Date
December 4, 2012
Termination Date
July 8, 2013
Address
16 Century Road, Neusoft Park,, Hun Nan Industrial Area, Shenyang, Liaoning, N/A, N/A, China

Description

NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

Reason

The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.

Code Info

NeuViz 16 system Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024. Neuviz Dual Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.

Distribution

Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.

Quantity

20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series