Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System
Recall: Z-0465-2013
·
Reported December 12, 2012
Enforcement
- Recall Number
- Z-0465-2013
- Event ID
- 63521
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 12, 2012
- Initiation Date
- June 20, 2012
- Classification Date
- December 3, 2012
- Termination Date
- January 14, 2014
- Address
- 51 Valley Stream Pkwy, N/A, Malvern, PA, 19355, United States
Description
Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System
Reason
Firm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename messages for the same patient are being continually processed for several days. This fills up the 'pcvdbsrv' log and potentially affects performance of the system. - Data could
Code Info
Model number 10 592 457
Distribution
Nationwide distribution: USA including states of: CA, FL, GA, MA, MN, MO, NJ, NY, OH, and WI.
Quantity
15