FDA Enforcement
Class II
Terminated
Affixus Hip Fracture Nail, RH 130 DEG 11MM X 280MM
Recall: Z-0463-2018
·
Reported February 7, 2018
Enforcement
- Recall Number
- Z-0463-2018
- Event ID
- 78763
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 7, 2018
- Initiation Date
- August 7, 2017
- Classification Date
- January 29, 2018
- Termination Date
- March 29, 2019
- Address
- 56 E Bell Dr, Warsaw, IN, 46582-6989, United States
Description
Affixus Hip Fracture Nail, RH 130 DEG 11MM X 280MM
Reason
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Code Info
Item No. 814511280 ; Lot No. (Exp. Date) UDI: 893430 (06/15/2027) UDI (01)00887868039126 (17)270615 (10)893430; 893490 (06/15/2027) UDI (01)00887868039126 (17)270615 (10)893490
Distribution
US nationwide distribution. Also Portugal and the Netherlands.
Quantity
638 units