FDA Enforcement Class II Ongoing

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment, Item Number: 720001B0

Recall: Z-0456-2025 · Reported November 20, 2024

Enforcement

Recall Number
Z-0456-2025
Event ID
95498
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Getinge Usa Sales Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 20, 2024
Initiation Date
October 3, 2024
Classification Date
November 14, 2024
Address
1 Geoffrey Way, N/A, Wayne, NJ, 07470-2035, United States

Description

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment, Item Number: 720001B0

Reason

Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

Code Info

UDI-DI: 04046768124332 S/N: 11399 through 11746

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.

Quantity

342 units