Merit Medical Cultura Collection and Transport System, 3 mL Viral Transport Medium, Ref: VCTS100, Single Use, IVD, for the following unit counts: 1 Unit - (01)00884450499917, 50 Units - (01)10884450499914, 200 units - (01)20884450499911
Recall: Z-0456-2021
·
Reported November 25, 2020
Enforcement
- Recall Number
- Z-0456-2021
- Event ID
- 86522
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Merit Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 25, 2020
- Initiation Date
- September 16, 2020
- Classification Date
- November 16, 2020
- Termination Date
- June 1, 2021
- Address
- 1600 W Merit Pkwy, N/A, South Jordan, UT, 84095-2416, United States
Description
Merit Medical Cultura Collection and Transport System, 3 mL Viral Transport Medium, Ref: VCTS100, Single Use, IVD, for the following unit counts: 1 Unit - (01)00884450499917, 50 Units - (01)10884450499914, 200 units - (01)20884450499911
Reason
The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in treatment.
Code Info
All Lots distributed prior to August 28, 2020. 1 Unit - (01)00884450499917, 50 Units - (01)10884450499914, 200 units - (01)20884450499911
Distribution
Worldwide distribution. US Nationwide, Australia, Canada, Colombia, Ghana, Ireland, Jamaica, Kenya, Panama, Qatar, Trinidad and Tobago.
Quantity
371,528 units