FDA Enforcement Class II Terminated

iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological

Recall: Z-0456-2014 · Reported December 18, 2013

Enforcement

Recall Number
Z-0456-2014
Event ID
66808
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Iba Dosimetry Gmbh
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
December 18, 2013
Initiation Date
August 20, 2013
Classification Date
December 6, 2013
Termination Date
February 24, 2014
Address
Bahnhofstr. 5, Schwarzenbruck, N/A, Germany

Description

iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological

Reason

The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.

Code Info

SW Version 3.0. Catalong Number CS10-100

Distribution

Worldwide Distribution - US Distribution including the states of CT, MA, OH, CA, SC, NY and OK, and the countries of International distribution (from Germany) Poland, Germany, GB, The Netherlands, Spain, Turkey, Canada, Mexico, China, Hong Kong and Singapore.

Quantity