FDA Enforcement
Class II
Terminated
iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological
Recall: Z-0456-2014
·
Reported December 18, 2013
Enforcement
- Recall Number
- Z-0456-2014
- Event ID
- 66808
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Iba Dosimetry Gmbh
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- December 18, 2013
- Initiation Date
- August 20, 2013
- Classification Date
- December 6, 2013
- Termination Date
- February 24, 2014
- Address
- Bahnhofstr. 5, Schwarzenbruck, N/A, Germany
Description
iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological
Reason
The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.
Code Info
SW Version 3.0. Catalong Number CS10-100
Distribution
Worldwide Distribution - US Distribution including the states of CT, MA, OH, CA, SC, NY and OK, and the countries of International distribution (from Germany) Poland, Germany, GB, The Netherlands, Spain, Turkey, Canada, Mexico, China, Hong Kong and Singapore.
Quantity
33 units