FDA Enforcement Class II Terminated

Merit Medical Cultura Collection and Transport System, Ref CFTS100, Single Use, IVD, for the following unit counts: 1 Unit - (01)00884450492123, 50 Units - (01)10884450492120, 200 units - (01)20884450492127

Recall: Z-0455-2021 · Reported November 25, 2020

Enforcement

Recall Number
Z-0455-2021
Event ID
86522
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Merit Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 25, 2020
Initiation Date
September 16, 2020
Classification Date
November 16, 2020
Termination Date
June 1, 2021
Address
1600 W Merit Pkwy, N/A, South Jordan, UT, 84095-2416, United States

Description

Merit Medical Cultura Collection and Transport System, Ref CFTS100, Single Use, IVD, for the following unit counts: 1 Unit - (01)00884450492123, 50 Units - (01)10884450492120, 200 units - (01)20884450492127

Reason

The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in treatment.

Code Info

All Lots distributed prior to August 28, 2020. 1 Unit - (01)00884450492123, 50 Units - (01)10884450492120, 200 units - (01)20884450492127

Distribution

Worldwide distribution. US Nationwide, Australia, Canada, Colombia, Ghana, Ireland, Jamaica, Kenya, Panama, Qatar, Trinidad and Tobago.

Quantity

159,396 units