FDA Enforcement
Class II
Terminated
Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a stand alone anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level from C2-T1.
Recall: Z-0455-2014
·
Reported December 11, 2013
Enforcement
- Recall Number
- Z-0455-2014
- Event ID
- 66538
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Synthes USA HQ, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 11, 2013
- Initiation Date
- September 30, 2013
- Classification Date
- December 5, 2013
- Termination Date
- April 16, 2015
- Address
- 1302 Wrights Ln E, N/A, West Chester, PA, 19380-3417, United States
Description
Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a stand alone anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level from C2-T1.
Reason
A potential non conformance was identified with the Zero-P VA plate where the screw could back out if the blocking mechanism is not visually confirmed to properly cover the head of the screw.
Code Info
part no. 04.647.117S, Plate implant no. 7706771, spacer lot no. 7675745
Distribution
Nationwide Distribution - only AZ .
Quantity
1