Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564
Enforcement
- Recall Number
- Z-0452-2022
- Event ID
- 89064
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomeme, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 12, 2022
- Initiation Date
- October 20, 2021
- Classification Date
- January 4, 2022
- Termination Date
- March 5, 2024
- Address
- 1015 Chestnut St, N/A, Philadelphia, PA, 19107-4316, United States
Description
Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564
The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).
All lots. Last lot of Kit distributed with the old IFU is 20210914.
Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA. Foreign distribution to Ghana, Dominican Republic, and Singapore.
11,607