FDA Enforcement Class II Terminated

Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564

Recall: Z-0452-2022 · Reported January 12, 2022

Enforcement

Recall Number
Z-0452-2022
Event ID
89064
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Biomeme, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 12, 2022
Initiation Date
October 20, 2021
Classification Date
January 4, 2022
Termination Date
March 5, 2024
Address
1015 Chestnut St, N/A, Philadelphia, PA, 19107-4316, United States

Description

Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564

Reason

The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).

Code Info

All lots. Last lot of Kit distributed with the old IFU is 20210914.

Distribution

Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA. Foreign distribution to Ghana, Dominican Republic, and Singapore.

Quantity

11,607