Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE
Enforcement
- Recall Number
- Z-0450-2026
- Event ID
- 97882
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- November 19, 2025
- Initiation Date
- October 10, 2025
- Classification Date
- November 13, 2025
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
GYN ROBOTIC DYNJ901046M UDI-DI 40198459130381 (case), 10198459130380 (ea) Lots 25FBM810 25GBK307; ROBOTIC PROSTATE DYNJ909120D UDI-DI 40198459130541 (case), 10198459130540 (ea) Lots 25GBT821 25IBF809
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
96,406 total units