Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING
Enforcement
- Recall Number
- Z-0448-2026
- Event ID
- 97882
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- November 19, 2025
- Initiation Date
- October 10, 2025
- Classification Date
- November 13, 2025
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
HIP PINNING DYNJ909308D UDI-DI 40198459055288 (case), 10198459055287 (ea) Lots 25FBB311 25GBC282 25GBW486; ORTHO SPINE ST CLAIR DYNJ901048N UDI-DI 40198459055554 (case), 10198459055553 (ea) Lots 25GBQ070; SET UP DYNJ909132F UDI-DI 40198459130572 (case), 10198459130571 (ea) Lots 25GBI845 25IBF708; UROLOGY MINOR DYNJ909123F UDI-DI 40198459130602 (case), 10198459130601 (ea) Lots 25GBT826 25IBF707
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
96,406 total units