FDA Enforcement Class II Ongoing

Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP

Recall: Z-0447-2022 · Reported January 12, 2022

Enforcement

Recall Number
Z-0447-2022
Event ID
89062
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Invacare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 12, 2022
Initiation Date
November 12, 2021
Classification Date
January 3, 2022
Address
1200 Taylor St, N/A, Elyria, OH, 44035-6248, United States

Description

Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP

Reason

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

Code Info

LiNX Gyro firmware version 6.1.2 UDI-DI Numbers: SRX-20MP 00841447114581 SRX-20R 00841447114604 SRX-20SP 00841447114598

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.

Quantity

1 unit US