FDA Enforcement
Class II
Ongoing
Computed Tomography X-Ray System, Model uCT 760, Rx, CE, GTIN: 06971576831043 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).
Recall: Z-0447-2021
·
Reported November 18, 2020
Enforcement
- Recall Number
- Z-0447-2021
- Event ID
- 86269
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 18, 2020
- Initiation Date
- August 5, 2020
- Classification Date
- November 12, 2020
- Address
- No. 2258 Chengbei Rd, Jiading Ind., Shanghai, N/A, China
Description
Computed Tomography X-Ray System, Model uCT 760, Rx, CE, GTIN: 06971576831043 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).
Reason
A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.
Code Info
Serial Numbers: 600229, 600230, 600232, 600271, 600282
Distribution
US Nationwide distribution including in the state of TX.
Quantity
5