FDA Enforcement Class II Ongoing

Computed Tomography X-Ray System, Model uCT 550, Rx, CE, GTIN: 06971576831036 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Recall: Z-0446-2021 · Reported November 18, 2020

Enforcement

Recall Number
Z-0446-2021
Event ID
86269
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Shanghai United Imaging Healthcare Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 18, 2020
Initiation Date
August 5, 2020
Classification Date
November 12, 2020
Address
No. 2258 Chengbei Rd, Jiading Ind., Shanghai, N/A, China

Description

Computed Tomography X-Ray System, Model uCT 550, Rx, CE, GTIN: 06971576831036 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Reason

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Code Info

Serial Numbers: 353020, 353021, 353023, 353045, 353049

Distribution

US Nationwide distribution including in the state of TX.

Quantity

5