Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
Enforcement
- Recall Number
- Z-0445-2026
- Event ID
- 97882
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- November 19, 2025
- Initiation Date
- October 10, 2025
- Classification Date
- November 13, 2025
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
ENT DYNJ909128I UDI-DI 40198459333355 (case), 10198459333354 (ea) Lots 25GBO685; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D UDI-DI 40198459055301 (case), 10198459055300 (ea) Lots 25GBN602
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
96,406 total units