FDA Enforcement Class I Ongoing

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F

Recall: Z-0444-2026 · Reported November 19, 2025

Enforcement

Recall Number
Z-0444-2026
Event ID
97882
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
November 19, 2025
Initiation Date
October 10, 2025
Classification Date
November 13, 2025
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F

Reason

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Code Info

DYNJ909116D UDI-DI 40198459055356 (case), 10198459055355 (ea) Lots 25FBM604 25GBP604; DYNJ909124F UDI-DI 40198459130619 (case), 10198459130618 (ea) Lots 25FBB831 25GBI833

Distribution

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Quantity

96,406 total units