FDA Enforcement Class II Ongoing

Catalys Precision Laser System, Models: Catalys-U (United States), Catalys-I (International), Catalys-C (China); and Catalys Precision Laser System No Bed, Models: 0160-6020 (United States), 0160-6010 (International)

Recall: Z-0444-2022 · Reported January 5, 2022

Enforcement

Recall Number
Z-0444-2022
Event ID
88986
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
AMO Manufacturing USA, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 5, 2022
Initiation Date
October 21, 2021
Classification Date
December 29, 2021
Address
510 Cottonwood Dr, Milpitas, CA, 95035-7403, United States

Description

Catalys Precision Laser System, Models: Catalys-U (United States), Catalys-I (International), Catalys-C (China); and Catalys Precision Laser System No Bed, Models: 0160-6020 (United States), 0160-6010 (International)

Reason

Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.

Code Info

Catalys-U and Catalys-I with Windows 7 Ultimate/Professional/Embedded Standard. Catalys-C with Windows 7 Embedded Standard. 0160-6020 and 0160-6010 with Windows 7 Professional/Embedded Standard.

Distribution

US: CA, TX, FL, LA, MD, IA, MA, VA, SD, MN, MO, NC, TN, SC, WA, NV, OH, KY, IN, CO, NJ, OK, NH, UT, AR, NY, AK, ND, NE, IL, AL, HI, CT, DE, KS, GA, OR, MS, PA, MT, ME, MI, AZ, ID, WI, VT, NM, CA. OUS: Canada, Mexico, Russian Federation, Turkey, Brazil, Egypt, Iraq, Saudi Arabia, Jordan, Norway, Colombia, Germany, Malaysia, India, Oman, Azerbaijan, Canada, Slovakia, Lebanon, Spain, Ireland, Italy, South Korea, Bangladesh, Chile, Taiwan, Tunisia, Kazakhstan, Libya, Argentina, Singapore, Austria, Australia, Vietnam, Greece, Poland, Indonesia, Yemen, Great Britain, South Africa, Croatia, France, Czech Republic, Algeria, Thailand, Bulgaria, United Arab Emirates, Belgium, Japan, China, Nepal, Israel, Switzerland, Dominican Republic, Hong Kong, Portugal, Ecuador, Kuwait, Netherlands, Slovenia, Philippines

Quantity

546