iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United States); iDesign Refractive Studio Aberrometer (International) and (United States), Model: G301
Enforcement
- Recall Number
- Z-0443-2022
- Event ID
- 88986
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- AMO Manufacturing USA, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 5, 2022
- Initiation Date
- October 21, 2021
- Classification Date
- December 29, 2021
- Address
- 510 Cottonwood Dr, Milpitas, CA, 95035-7403, United States
Description
iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United States); iDesign Refractive Studio Aberrometer (International) and (United States), Model: G301
Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.
G300 with Windows XP. G301 with Windows 7 Embedded System.
US: CA, TX, FL, LA, MD, IA, MA, VA, SD, MN, MO, NC, TN, SC, WA, NV, OH, KY, IN, CO, NJ, OK, NH, UT, AR, NY, AK, ND, NE, IL, AL, HI, CT, DE, KS, GA, OR, MS, PA, MT, ME, MI, AZ, ID, WI, VT, NM, CA. OUS: Canada, Mexico, Russian Federation, Turkey, Brazil, Egypt, Iraq, Saudi Arabia, Jordan, Norway, Colombia, Germany, Malaysia, India, Oman, Azerbaijan, Canada, Slovakia, Lebanon, Spain, Ireland, Italy, South Korea, Bangladesh, Chile, Taiwan, Tunisia, Kazakhstan, Libya, Argentina, Singapore, Austria, Australia, Vietnam, Greece, Poland, Indonesia, Yemen, Great Britain, South Africa, Croatia, France, Czech Republic, Algeria, Thailand, Bulgaria, United Arab Emirates, Belgium, Japan, China, Nepal, Israel, Switzerland, Dominican Republic, Hong Kong, Portugal, Ecuador, Kuwait, Netherlands, Slovenia, Philippines
711