FDA Enforcement Class II Terminated

Devices includes a Multi Port Inflator or without Multi Port Inflator (MPI). All the devices are vascular blood flow machines. Devices are distributed with pressure cuffs of varying sizes as accessories. Affected models include *2100-SX, 3000 series, 2017, 2016, 2015, 2014, 1059, 1059-A, *1059-C, *1058-C, 4000 series, 1100 series, 1060 series, 1080 series, 234 and 246. (* = current production model) All the devices are used in the study and diagnosis of vascular disease and blockages in the upper and lower extremities.

Recall: Z-0443-2013 · Reported December 5, 2012

Enforcement

Recall Number
Z-0443-2013
Event ID
63235
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Parks Medical Electronics, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
December 5, 2012
Initiation Date
August 8, 2012
Classification Date
November 27, 2012
Termination Date
August 19, 2016
Address
19460 SW Shaw St, N/A, Aloha, OR, 97007-1242, United States

Description

Devices includes a Multi Port Inflator or without Multi Port Inflator (MPI). All the devices are vascular blood flow machines. Devices are distributed with pressure cuffs of varying sizes as accessories. Affected models include *2100-SX, 3000 series, 2017, 2016, 2015, 2014, 1059, 1059-A, *1059-C, *1058-C, 4000 series, 1100 series, 1060 series, 1080 series, 234 and 246. (* = current production model) All the devices are used in the study and diagnosis of vascular disease and blockages in the upper and lower extremities.

Reason

Parks Medical Electronics, Inc. was made aware of two incidents where an air supply tube from their PARKS PVR unit was inadvertently connected to an IV in the patient's arm instead of into the blood pressure cuff.

Code Info

Domestic Distribution: 013053, 013104, 013107, 013124, 013154, 013182, 013193, 014203, 014237, 014248, 014249, 014252, 014253, 014254, 014269, 014276, 014310, 014335, 014339, 014361, 014366, 014376, 014380, 014381, 014382, 014384, 014397, 014401, 014406, 014416, 014424, 014429, 014431, 014443, 014444, 014448, 014454, 014458, 014530, 014532, 014533, 014540, 014542, 014545, 014547, 014551, 014552, 014552, 014553, 014598, 014601, 014604, 014608, 014616, 014654, 014657, 014659, 014660, 014661, 014663, 103077, 105X-600001024, 105X-600001025, 105X-600001026, 105X-600001027, 105X-800001019, 105X-800001022, 105X-800001023, 105X-800001024, 105X-800001026, 105X-800001027, 105X-800001027, 105X-800001028, 105X-800001029, 105X-800001030, 105X-800001031, 105X-E00001001, 105X-E00001002, 105X-E00001003, 105X-E00001004, 105X-E00001005, 105X-E00001007, 105X-E00001008, 105X-E00001009, 105X-E00001010, 105X-E00001011, 105X-E00001012, 105X-E00001013, 105X-E00001014, 105X-E00001016, 105X-E00001017, 105X-E00001018, 105X-E00001019, 105X-E00001020, 105X-F00001000, 105X-F00001001, 105X-F00001002, 105X-F00001003, 105X-F00001004, 105X-F00001005, 105X-F00001005, 105X-F00001006, 105X-F00001007, 105X-F00001008, 105X-F00001009, 105X-F00001010, 105X-F00001011, 105X-F00001012, 105X-F00001013, 105X-F00001014, 105X-F00001015, 105X-F00001016, 105X-F00001017, 105X-F00001018, 105X-F00001020, 105X-F00001021, 105X-F00001022, 105X-F00001023. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify. Foreign distribution: 01301, 013043, 013044, 013045, 013046, 013072, 013076, 013175, 013181, 013183, 013188, 014215, 014386, 014392, 014393, 014456, 014535, 014550, 014551, 014561, 014562, 014563, 105X-800001025, 105X-922001017, 105X-922001018, 105X-922001019, 105X-922001021, 105X-9221020, 105X-923001001, 105X-923001002, 105X-923001003, 105X-923001004, 105X-923001005, 105X-923001006, 105X-923001007, 105X-923001007, 105X-923001009, 105X-923001010, 105X-923001011, 105X-E00001000, 105X-E00001003, 105X-E00001006, 105X-E00001015, 105X-F00001019. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify.

Distribution

Worldwide distribution: USA (nationwide) and countries of: Australia, Brazil, Canada, Chile, Egypt, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Korea, Mexico, Peru, Republica Argentina, Spain, Switzerland, and Venezuela.

Quantity

349 units (Domestic: 270 units - Foreign: 79 units)