FDA Enforcement
Class II
Terminated
v|tome|x L Series non-destructive testing cabinet x-ray systems
Recall: Z-0442-2021
·
Reported November 25, 2020
Enforcement
- Recall Number
- Z-0442-2021
- Event ID
- 86620
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ge Inspection Technologies Llc
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- Letter
- Report Date
- November 25, 2020
- Initiation Date
- September 29, 2020
- Classification Date
- November 13, 2020
- Termination Date
- May 2, 2023
- Address
- 721 Visions Dr, N/A, Skaneateles Falls, NY, 13153-5314, United States
Description
v|tome|x L Series non-destructive testing cabinet x-ray systems
Reason
It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on.
Code Info
v/tome/x L Series (L240, L300 or L450)
Distribution
USA Canada
Quantity
N/A