FDA Enforcement Class II Ongoing

bk5000 Ultrasound System w/battery; Model No. 2300-61; System, Imaging, Pulsed Doppler, Ultrasonic

Recall: Z-0440-2026 · Reported November 12, 2025

Enforcement

Recall Number
Z-0440-2026
Event ID
97830
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
B-K Medical A/S
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 12, 2025
Initiation Date
October 9, 2025
Classification Date
November 6, 2025
Address
Mileparken 34, N/A, Herlev, N/A, N/A, Denmark

Description

bk5000 Ultrasound System w/battery; Model No. 2300-61; System, Imaging, Pulsed Doppler, Ultrasonic

Reason

GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.

Code Info

GTIN 05704916000264 and pending, Serial Numbers: 2002509, 2008698, 2006429, 5013043, 2008314, 2002617, 5016133, 5013054, 2005257, 2003097, 2003133, 2002655, 2001908, 2001567, 2001587, 5014656, 2001346, 2003432, 5015936, 5015937, 2007537, 2007493, 2007508, 2005573, 2003569, 2005652, 2002618, 5015938, 5015966, 2007534, 2005979, 2002510, 2003079, 2003083, 5013847, 2003532, 2003535, 5015614, 5015616, 5015617, 5015677, 5015720, 5015721, 5009669, 2003562, 5015927, 2001804, 5015611, 2008819, 2004819, 2004816, 2004939, 5016388, 5016390, 5009259, 5016395, 5015600, 5014860, 5008890, 5016164, 2000873, 2000878, 2000118, 2000880, 2001884, 5008010, 2003034, 5014063, 2006387, 5015116, 5010745, 5014406, 5014407, 5014408, 2008604, 5010979, 5010322, 5010408, 2003564, 2006191, 5014975, 2006407, 2005904, 2005920, 2006218, 5013389, 5011535, 2005357, 5011225, 2006894, 2005844, 2003526, 2003568, 5009258, 2005666, 2003923, 2009162, 2009149, 2008912, 2004177, 2004782, 2006039, 2005657, 2001916, 2001917, 5014174, 5008020, 2000876, 5015684, 5009371, 5009665, 2007644, 5011701, 5013040, 5014121, 5014655, 5014657, 5015225, 5015228, 5015924, 5015925, 2002652, 5012534, 2006422, 2007460, 2002405, 2001515, 2001516, 2004114, 2004438, 2008630, 2008811, 2003528, 2007507, 5015986, 2004121, 5014351, 2003434, 2006071, 2007573, 2006657, 2007540, 5016233, 5013794, 2001565, 2008024, 2008324, 5014992, 5016448, 2007466, 5015363, 2005816, 2005957, 2005964, 2006769, 2006829, 2001119, 2001122, 2003229, 2006112, 2006139, 2006203, 2005766, 2007346, 5016222, 2007602, 2009217, 2006457, 2008021, 2008178, 2002227, 2002229, 2002236, 2002237, 5015968, 5007860, 5007902, 5007909, 2001913, 5013051, 5013053, 5016235, 5011226, 2002649, 2003979, 2001583, 5010417, 2003081, 5013232, 5016491, 5016492, 2001123, 2003567, 2003570, 5009254, 5013770, 5015487, 5015491, 5009682, 5016185, 2002103, 2004052, 5016040, 2008930, 2001882, 2001909, 2002244, 5016418, 2003232, 2003259, 2005175, 2006402, 2006441, 2007343, 2007345, 2004840, 5016234, 5016236, 5016240, 5016336, 2002234, 5011170, 5015224, 2003098, 5016490, 5009256, 2003321, 2004594, 2003317, 2006159, 2001559, 2001566, 2001586, 2003563, 2005391, 5013027, 2004662, 2001513, 2006898, 2003101, 2007643, 2002243, 2002560, 2006746, 2008016, 2008706, 2003262, 2006409, 2002240, 2002561, 2002619, 2002624, 2006181, 2008012, 5010412, 5014507, 5015965, 2004690, 5012509, 5012513, 5013340, 5013341, 5009379, 5015686, 5015687, 5013587, 5013589, 5010705, 5013796, 2005762, 2008264, 2008291, 2000792, 2000793, 2000794, 2000795, 2000797, 2000961, 2000967, 2000970, 2000725, 2002238, 2001285, 2008890, 5014730, 5016029, 2000958, 2002111, 2004813, 5014506, 2002241, 2003323, 5013390, 2001449, 2006477, 2002504, 5015722, 5016172, 2007474, 2003530, 2008641, 2001802, 5012455, 5008837, 5008886, 2003258, 2003130, 5014247, 5016396, 5016402, 5016403, 5016404, 5013848, 5015496, 5012852, 5012854, 5014350, 5015230, 5015250, 2002734, 2001560, 2001562, 2001563, 5008008, 2004496, 5016182, 2003982, 5008771, 5010418, 2006390, 2002109, 2004208, 2004678, 5016344, 2006301, 5016041, 5015967, 2007341, 5016183, 2008823, 2004808, 2007458, 5008885, 5008887, 2007532, 5016339, 5010982, 5009370, 5010976, 2008583, 2008615, 5012851, 5009496, 5016337, 2004188, 5013316, 5016414, 5016223, 5015497, 5015612, 5013916, 5013917, 2000872, 2004526, 2005448, 2002164, 2005847, 5014829, 2000955, 2000722, 2001582, 2000611, 2001116, 2000721, 2001347, 2003231, 2000874, 5016242, 5016243, 2000614, 2003319, 2000119, 5009679, 5014374, 2000660, 2004523, 5015227, 5014234, 2008610, 2008632, 2008853, 5009681, 5015493, 5009509, 2007323, 2007480, 2007487, 5011167, 2001130, 5014843, 5007892, 5011250, 5008005, 5008017, 5008668, 2006350, 2002505, 2002955, 2003233, 2003430, 2003431, 2003905, 2003985, 2004542, 2004760, 2004775, 2004907, 2005921, 2003263, 2003320, 2003529, 2003318, 2002622, 2004911, 2007715, 2001878, 2001880, 2001883, 2001886, 2004205, 2008924, 2008177, 2008004, 2008022, 2005376, 5014375, 2005550, 2005274, 2001512, 2006017, 2004085, 2000881, 2003256, 2005582, 2005753, 2003884, 5012528, 2003322, 5015045, 5015108, 5012079, 5012163, 2001799, 5011321, 5016126, 5016413, 5016427, 2004815, 2003920, 2003960, 2005538, 2005545, 5015399, 2001127, 2001564, 2006335, 2001348, 5014658, 5013583, 2005423, 2000664, 5016383, 2000799, 2000800, 2006106, 5015226, 2000618, 2000965, 2008961, 2001124, 2001286, 2001806, 2004776, 2004821, 2005643, 5012252, 5011929, 2006187, 5009582, 2000971, 2000728, 2000730, 5008896, 5016167, 5015046, 5012665, 5011333, 2000619, 2006482, 2002621, 2001578, 5011930, 2006463, 2001885, 2003260, 5013615, 5011330, 2004645, 2000729, 2002939, 2005808, 2004661, 2006419, 2001457, 2000616, 5013356, 5010997, 5014064, 2004924, 2000724, 2004569, 2001352, 5015257, 2002653, 2003435, 5011000, 5014176, 2000668, 5008894, 5011927, 5013630, 5015982, 2004502, 5015607, 5015608, 5015403, 5014991, 5014859, 2006531, 5013357, 5012077, 2002107, 2004115, 5009660, 2003428, 5010211, 2004029, 2009209, 5015486, 5013248, 5015113, 2006836, 5016121, 5013631, 5015402, 5013776, 5015110, 2003226, 5015038, 2006466, 5015689, 2004666, 2001125, 5009373, 2004806, 5016105, 5014846, 2000617, 5008013, 2007500, 5012590, 2007510, 5014787, 2004607, 2004669, 2001290, 5008884, 5015361, 2001292, 2001580, 2002961, 5008883, 2001113, 2000791, 2006190, 2003436, 2004818, 2004914, 2005162, 2000610, 5011829, 5011836, 5011326, 2006781, 5010998, 5010999, 5015613, 2000966, 5007751, 5010698, 5014505, 5013846, 2006565, 2007494, 2001801, 5015375, 5015609, 2004856, 2004916, 5014852, 5016263, 2004110, 5008002, 5015984, 2003104, 5014108, 2004076, 5009683, 5014175, 2000669, 5014177, 2003994, 2007471, 2007543, 2008015, 2008311, 2008322, 2008331, 2008344, 2008405, 2008432, 2008619, 2008627, 2008699, 2008860, 2008865, 2008926, 5012662, 5015604, 5014243, 5015400, 2002648, 2000663, 5016276, 2003135, 5016326, 5016327, 2000726, 5009372, 2004011, 2005408, 5013616, 2002737, 2005006, 5015039, 2000877, 5009670, 2001510, 2004027, 2004034, 2001805, 2004676, 5015964, 5015685, 5015043, 5016226, 2002362, 2007574, 2006042, 5010992, 2008342, 2008596, 2003132, 5013380, 5011533, 2006675, 2001511, 2000953, 2002956, 2003103, 2003126, 2003576, 2006544, 2007366, 5013230, 5015610, 5015934, 5016184, 2001452, 2004079, 2004932, 5016039, 5016221, 5009595, 5010210, 5007895, 2003566, 2000661, 5016224, 5013033, 5014246, 2002964, 5013038, 5016266, 2004778, 5011324, 5013614, 5008195, 5014289, 2000727, 5014178, 5016335, 5015413, 5015414, 2000875, 5015602, 5012859, 5016288, 5007063, 5014109, 5015601, 2003534, 2003325, 5016106, 2005937, 2001289, 2001581, 2002226, 2002741, 2002743, 2003889, 2004003, 2004196, 2004759, 2004837, 5009247, 5009658, 5010776, 5012532, 5013777, 5014245, 5014659, 5014976, 5015485, 5015488, 5015490, 5015926, 5016289, 2002647, 2001454, 5009672, 5013582, 5015401, 5012855, 2008939, 5015251, 5015255, 5015256, 5015258, 2004570, 5015407, 5012652, 5012600, 2000662, 2002411, 2004677, 2008644, 2000665, 2001912, 2002408, 2002958, 2003977, 2004043, 2005586, 2005806, 2005877, 2007813, 5013795, 2000723, 2000964, 2002108, 2002231, 2002625, 2005349, 2007465, 2008392, 2008403, 5016137, 5016165, 5016166, 2001117, 2004087, 2004436, 2004606, 2004659, 2004844, 2004915, 2004918, 2006013, 2006031, 2006083, 2006086, 2006321, 2006376, 2006399, 2006412, 2006444, 2006447, 2006452, 2006461, 2007586, 2007866, 2008599, 2008270, 2008260, 2008608, 2008875, 2008956, 2008304, 2008293, 2008305, 2008317, 2008329, 2000116, 2000968, 2000969, 2001351, 2002736, 2003227, 2004213, 2004384, 2005560, 2006040, 2008020, 5009364, 5010419, 5014778, 5015378, 5016227, 2007327, 2007364, 2007469, 2007472, 2002105, 2002513, 2004475, 2008442, 5013320, 5013321, 2002740, 2001287, 2001293, 2001294, 2001585, 2002239, 2004086, 5013782, 5012247, 5012249, 2005789, 2005016, 5010985, 5010989, 5010410, 2008818, 2001910, 2001121, 2001129, 2003316, 2005370, 2005647, 2006799, 2002940, 2005552, 2007639, 2008895, 5014788, 5016244, 2004002, 2005415, 5010747, 5011322, 5014725, 5014726, 5012927, 2004934, 5008189, 2005454, 2007344, 2006027, 2006748, 2008028, 2008398, 5009253, 2006016.

Distribution

Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkiye, Ukraine, United Arab Emirates, and United Kingdom.

Quantity

895 units