FDA Enforcement
Class II
Ongoing
VIDAS SARS-COV-2 IgG, Ref 423834-01
Recall: Z-0439-2022
·
Reported January 5, 2022
Enforcement
- Recall Number
- Z-0439-2022
- Event ID
- 88797
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- bioMerieux, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- January 5, 2022
- Initiation Date
- September 22, 2021
- Classification Date
- December 28, 2021
- Address
- 100 Rodolphe St, Durham, NC, 27712-9402, United States
Description
VIDAS SARS-COV-2 IgG, Ref 423834-01
Reason
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.
Code Info
UDI 03573026620745, Lot #: 1008671430, 1008763910
Distribution
US Nationwide distribution.
Quantity
Two lots