FDA Enforcement Class II Terminated

Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiographic techniques.

Recall: Z-0439-2020 · Reported November 27, 2019

Enforcement

Recall Number
Z-0439-2020
Event ID
84125
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cook Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 27, 2019
Initiation Date
October 21, 2019
Classification Date
November 18, 2019
Termination Date
June 30, 2020
Address
750 N Daniels Way, N/A, Bloomington, IN, 47404-9120, United States

Description

Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiographic techniques.

Reason

The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.

Code Info

REF (RPN): HNB5.0-38-100-P-NS-VERT REF (GPN): G06074 Lot Numbers: 9696580, 9707470, 9707471, 9710400, 9710402, 9710403, 9710404, 9710405, 9734099, NS9710401 UDIs: (01)00827002060746(17)220508(10)9696580, (01)00827002060746(17)220513(10)9707470, (01)00827002060746(17)220514(10)9707471, (01)00827002060746(17)220515(10)9710400, (01)00827002060746(17)220516(10)9710402, (01)00827002060746(17)220515(10)9710403, (01)00827002060746(17)220515(10)9710404, (01)00827002060746(17)220514(10)9710405, (01)00827002060746(17)220524(10)9734099, (01)00827002060746(17)220513(10)NS9710401

Distribution

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, IL, IN, KS, LA, MA, MD, Ml, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and PR. Countries of AU, BE, CA, CN, CZ, DK, FR, DE, GT, HK, IT, KR, LU, MX, NL, PK, PL, ZA, ES, SE, CH, TW, and GB.

Quantity

740