FDA Enforcement
Class II
Terminated
IO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-Post Reamer is an accessory to the IO FiX Screw and Washer System that is used for reaming bone most common to orthopedic surgical procedures.
Recall: Z-0438-2013
·
Reported December 5, 2012
Enforcement
- Recall Number
- Z-0438-2013
- Event ID
- 62918
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Extremity Medical LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 5, 2012
- Initiation Date
- April 25, 2011
- Classification Date
- November 26, 2012
- Termination Date
- March 21, 2013
- Address
- 300 Interpace Pkwy Ste 410A, N/A, Parsippany, NJ, 07054-1148, United States
Description
IO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-Post Reamer is an accessory to the IO FiX Screw and Washer System that is used for reaming bone most common to orthopedic surgical procedures.
Reason
Product Event Reports revealed a broken reamer experienced intra-operatively. Pre-drilling was not performed prior to reaming.
Code Info
Product coding is listed as Catalog Number: Lot Number and are as follows: 118-0005:126564; 118-30016:127375; 118-30018:AP6A08; 118-30020:AP6A07.
Distribution
Worldwide Distribution - USA including FL, MA, NC, CA, MD, VA, IN, AL, NJ, AZ, OK, MT, NY, OH, IA, MI, TX, WI, GA, IL, TN and NE. Internationally to Switzerland.
Quantity
156 units