FDA Enforcement Class II Ongoing

VIDAS Lyme IgG II (LYM), REF 417401

Recall: Z-0437-2022 · Reported January 5, 2022

Enforcement

Recall Number
Z-0437-2022
Event ID
88797
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
bioMerieux, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
January 5, 2022
Initiation Date
September 22, 2021
Classification Date
December 28, 2021
Address
100 Rodolphe St, Durham, NC, 27712-9402, United States

Description

VIDAS Lyme IgG II (LYM), REF 417401

Reason

bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.

Code Info

UDI 03573026458270, Lot #: 1008714880, 1008824220

Distribution

US Nationwide distribution.

Quantity

Two lots