FDA Enforcement Class II Terminated

MiniMed 620G, 3.0mL mg/dL, Model No. MMT-1710 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous monitoring of glucose levels in the fluid under the skin, and possible low and high blood glucose episodes.

Recall: Z-0437-2016 · Reported December 23, 2015

Enforcement

Recall Number
Z-0437-2016
Event ID
72352
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medtronic MiniMed Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 23, 2015
Initiation Date
September 25, 2015
Classification Date
December 17, 2015
Termination Date
February 5, 2016
Address
18000 Devonshire St, N/A, Northridge, CA, 91325-1219, United States

Description

MiniMed 620G, 3.0mL mg/dL, Model No. MMT-1710 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous monitoring of glucose levels in the fluid under the skin, and possible low and high blood glucose episodes.

Reason

The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.

Code Info

NG1019903H NG1019906H NG1019907H NG1019909H NG1019914H NG1019915H NG1019916H NG1019918H NG1019921H NG1019922H NG1019923H NG1019925H NG1019926H NG1019928H NG1019930H NG1019931H NG1019932H NG1019933H NG1019935H NG1019937H NG1019939H NG1019940H NG1019943H NG1019944H NG1019946H NG1019949H NG1019950H NG1019953H NG1019966H NG1019971H NG1019975H NG1019976H NG1019979H NG1019981H NG1019989H NG1019997H NG1019999H NG1020001H NG1020006H NG1020007H NG1020012H NG1020015H NG1020017H NG1020018H NG1020019H NG1020020H NG1020022H NG1020023H NG1020026H NG1020027H NG1020028H NG1020029H NG1020031H

Distribution

Internationally Distribution only to the countries of : Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya, Luxembourg, Namibia, Netherlands, Norway, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom.

Quantity

53 units