FDA Enforcement Class II Terminated

SIEMENS brand ONCOR series Linear Accelerator Linac systems with component: any photon unflat beam option(Multiple X feature) in combination with the IMRT or mARC option. A family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The ARTISTE/ONCOR systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, GERMANY and distributed SIEMENS Healthcare IM CR RO, CR  Radiation Oncology, Martinez, CA.

Recall: Z-0437-2014 · Reported December 11, 2013

Enforcement

Recall Number
Z-0437-2014
Event ID
66724
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 11, 2013
Initiation Date
October 23, 2013
Classification Date
December 3, 2013
Termination Date
January 22, 2014
Address
757 Arnold Dr Ste A, N/A, Martinez, CA, 94553-3615, United States

Description

SIEMENS brand ONCOR series Linear Accelerator Linac systems with component: any photon unflat beam option(Multiple X feature) in combination with the IMRT or mARC option. A family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The ARTISTE/ONCOR systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, GERMANY and distributed SIEMENS Healthcare IM CR RO, CR  Radiation Oncology, Martinez, CA.

Reason

A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.

Code Info

Serial Numbers: 5281, 5385, 5445

Distribution

Worldwide Distribution: USA (nationwide) including states of: FL, NC, NY, PA, TN, UT, and WI; and countries of: Angola, Brazil, Canada, P.R. China, Germany, India, Japan, Republic Korea, Poland, and Thailand.

Quantity

3 units