FDA Enforcement Class II Terminated

UniCel DxH 600 Coulter Cellular Analysis System, Software Version 3.0.2.0, Part No. 775222 The UniCel DxH 800/DxH 600 analyzers are quantitative, multi-parameter, automated hematology analyzers for in vitro diagnostic use in screening patient populations found in clinical laboratories. The DxH 800 is available as an analyzer for use on a benchtop or with a floor stand. The DxH 600 is available as an analyzer for use on a benchtop only.

Recall: Z-0433-2016 · Reported December 23, 2015

Enforcement

Recall Number
Z-0433-2016
Event ID
72773
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 23, 2015
Initiation Date
December 1, 2015
Classification Date
December 16, 2015
Termination Date
January 20, 2017
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

UniCel DxH 600 Coulter Cellular Analysis System, Software Version 3.0.2.0, Part No. 775222 The UniCel DxH 800/DxH 600 analyzers are quantitative, multi-parameter, automated hematology analyzers for in vitro diagnostic use in screening patient populations found in clinical laboratories. The DxH 800 is available as an analyzer for use on a benchtop or with a floor stand. The DxH 600 is available as an analyzer for use on a benchtop only.

Reason

Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but different Patient identification when manually editing pending orders at the System Manager.

Code Info

all serial numbers Part No. 775222

Distribution

Worldwide Distribution - US (nationwide) and Internationally to Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brunei Darussalam, Canada, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Lebanon, Lithuania, Macao, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Norway, Oman, Panama Philippines, Poland, Portugal, Puerto Rico , Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Republic, United Kingdom, Uruguay, and Vietnam.

Quantity

834 units total (594 units in US)