syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R) Automation Solutions. It provides autoverification, integrated QC capabilities, browser-based connectivity supporting up to 25 concurrent users, remote services to help keep systems running optimally and also has audit trail capabilities.
Enforcement
- Recall Number
- Z-0433-2014
- Event ID
- 65147
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 11, 2013
- Initiation Date
- May 10, 2013
- Classification Date
- December 3, 2013
- Termination Date
- August 26, 2014
- Address
- 500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States
Description
syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R) Automation Solutions. It provides autoverification, integrated QC capabilities, browser-based connectivity supporting up to 25 concurrent users, remote services to help keep systems running optimally and also has audit trail capabilities.
Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.
Siemens Material Number 10800057 (Version VA11B) and 10803188 (Version VA12A)
Worldwide Distribution: USA (nationwide) and internationally to Canada.
262