FDA Enforcement Class II Terminated

MYLA MASTER DVD V4.1 CLI Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).

Recall: Z-0429-2016 · Reported December 23, 2015

Enforcement

Recall Number
Z-0429-2016
Event ID
72643
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
bioMerieux, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 23, 2015
Initiation Date
November 3, 2015
Classification Date
December 16, 2015
Termination Date
December 4, 2017
Address
100 Rodolphe St, N/A, Durham, NC, 27712-9402, United States

Description

MYLA MASTER DVD V4.1 CLI Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).

Reason

MYLA¿ software connected to a Laboratory Information System (LIS) and a VITEK¿ 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.

Code Info

REF # 417191, Serial number/Software version V4.1 CLI

Distribution

Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and countries of: Foreign accounts: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, Germany, Greece, Hungary, India, Italy, Japan, Korea, Netherlands, Poland, Portugal, Russia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom, Angola, Puerto Rico, Republic of Belarus, Republic of Belarus, Benin, Bosnia, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Chad, Democratic Republic of Congo, Costa Rica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Gabon, Guam, Guatemala, Guatemala, Guinea, Honduras, Indonesia, Iraq, Israel, Ivory Coast, Kazakhstan, Kenya, Kenya, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Malaysia, Malaysia, Mali, Portugal, Myanmar, Managua, Niger, Nigeria, Palestine, Philippines, Philippines, Qatar, Kingdom of Saudi Arabia, Senegal, Serbia, Slovakia, Togo, Tunisia, Ukraine, UAE, Vietnam, Vietnam, Ivory Coast and Zimbabwe.

Quantity

626 systems in total