FDA Enforcement Class II Ongoing

Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer

Recall: Z-0428-2026 · Reported November 12, 2025

Enforcement

Recall Number
Z-0428-2026
Event ID
97730
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
November 12, 2025
Initiation Date
October 1, 2025
Classification Date
November 4, 2025
Address
1000 Lake Hazeltine Dr, N/A, Chaska, MN, 55318-1037, United States

Description

Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer

Reason

An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.

Code Info

A71460, UDI/DI 15099590369224, Serial Numbers: 902515, 902529, 902528, 902534, 902526, 902527, 902523, 902522, 902524, 902505, 902578, 902579, 902545, 902530, 902531, 902581, 902504, 902514, 902525, 902535, 902566, 902516, 902518, 902532, 902555, 902558, 902564, 902519, 902521, 902559, 902511, 902549, 902554, 902547, 902582, 902583; A30260, UDI/DI 15099590341602, Serial Numbers: 902536, 902541, 902517, 902563, 902565, 902557, 902513, 902507, 902543, 902560, 902562, 902506, 902508, 902574, 902533, 902572, 902540, 902569, 902568, 902577, 902510, 902580, 902567, 902573, 902570, 902571, 902538, 902553, 902552, 902512, 902539, 902561, 902575, 902544, 902542, 902556, 902537, 902509, 902520, 902546, 902548, 902551, 902550, 902576.

Distribution

Worldwide

Quantity

80 units