FDA Enforcement
Class II
Terminated
13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.
Recall: Z-0425-2014
·
Reported December 4, 2013
Enforcement
- Recall Number
- Z-0425-2014
- Event ID
- 65356
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- B. Braun Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 4, 2013
- Initiation Date
- April 4, 2013
- Classification Date
- November 27, 2013
- Termination Date
- August 26, 2014
- Address
- 901 Marcon Boulevard, N/A, Allentown, PA, 18109-9512, United States
Description
13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.
Reason
Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i
Code Info
Material BC1300, Lot number 61286479.
Distribution
Nationwide Distribution
Quantity
4,000 units