FDA Enforcement Class II Terminated

13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.

Recall: Z-0425-2014 · Reported December 4, 2013

Enforcement

Recall Number
Z-0425-2014
Event ID
65356
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
B. Braun Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 4, 2013
Initiation Date
April 4, 2013
Classification Date
November 27, 2013
Termination Date
August 26, 2014
Address
901 Marcon Boulevard, N/A, Allentown, PA, 18109-9512, United States

Description

13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.

Reason

Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i

Code Info

Material BC1300, Lot number 61286479.

Distribution

Nationwide Distribution

Quantity

4,000 units