FDA Enforcement Class II Terminated

SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty.

Recall: Z-0423-2014 · Reported December 4, 2013

Enforcement

Recall Number
Z-0423-2014
Event ID
66864
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
DePuy Orthopaedics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
December 4, 2013
Initiation Date
November 18, 2013
Classification Date
November 27, 2013
Termination Date
February 24, 2015
Address
700 Orthopaedic Dr, N/A, Warsaw, IN, 46582-3994, United States

Description

SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT PUNCH NONCEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty.

Reason

DePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.

Code Info

Item #950502021

Distribution

Worldwide Distribution - USA (nationwide) including AR,AZ,CA,CO,FL,GA,HI,IA,IL,IN,KY,LA,MA,MD,ME,MI,MN,MT,NC,NV,NY,OH,PA, TX,VA,WA,WI and Internationally to: AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CZECH REPUBLIC DENMARK EGYPT FRANCE GERMANY GREECE HOLLAND HUNGARY INDIA IRELAND ISRAEL ITALY JAPAN KOREA MALAYSIA NEW ZEALAND NORWAY P R CHINA PAKISTAN PANAMA POLAND PORTUGAL PUERTO RICO RUSSIA SAUDI ARABIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TURKEY UAE UK

Quantity

29,235 units in total / 2008 through November 2013