FDA Enforcement Class I Ongoing

CARDIOHELP Emergency Drive, Part Numbers 701048002 and 701076205. The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system.

Recall: Z-0420-2024 · Reported December 13, 2023

Enforcement

Recall Number
Z-0420-2024
Event ID
93197
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Maquet Medical Systems USA
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 13, 2023
Initiation Date
October 10, 2023
Classification Date
December 7, 2023
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094, United States

Description

CARDIOHELP Emergency Drive, Part Numbers 701048002 and 701076205. The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system.

Reason

Product removal due to possible blocking or impairment of the CARDIOHELP Emergency Drive. This issue can lead to the patient being exposed to inappropriate low blood flow. Potential associated harms are ischemia and hypoxia.

Code Info

Part No. 701048002: UDI - 04037691643526; Serial Numbers 90425259 to 90425748 (Excluding 90425438, 90425443, 90425568, 90425570, 90425588, 90425716) [***Updated 1/8/24 to include Serial numbers 90421507, 90421980, 90424277. Serial numbers 90425366 and 90423121 are additionally included in the scope due to internal complaints, but these units were corrected prior to the field action initiation and therefore are not included in customer notification***] Part No. 701076205: No UDI; Serial Numbers 90425438, 90425443, 90425568, 90425570, 90425588, 90425716, 90425747, 90425748

Distribution

US Nationwide - Worldwide Distribution: Domestic distribution to: AL, CA, DC, FL, HI, IA, IL, IN, KY, MD, MI, MN, MO, NE, OH, SC, VA Foreign distribution to: Austria, Australia, Brazil, Canada, Chile, China, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Japan, Jordan, Mexico, New Zealand, Pakistan, Portugal, Russia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom

Quantity

Updated 1/8/24 - 1858 Total distributed in U.S.; 69 Total affected units in U.S.