FDA Enforcement
Class II
Terminated
MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM
Recall: Z-0419-2018
·
Reported February 7, 2018
Enforcement
- Recall Number
- Z-0419-2018
- Event ID
- 78886
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Medical Division of Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 7, 2018
- Initiation Date
- July 21, 2017
- Classification Date
- January 27, 2018
- Termination Date
- March 22, 2018
- Address
- 3800 E Centre Ave, Portage, MI, 49002-5826, United States
Description
MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM
Reason
It was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied with the reflective side toward the patient (upside down).
Code Info
LOT #(s): 1640002, 1645003
Distribution
U.S.
Quantity
11160