FDA Enforcement
Class II
Ongoing
GIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Recall: Z-0416-2024
·
Reported December 6, 2023
Enforcement
- Recall Number
- Z-0416-2024
- Event ID
- 93335
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Aizu Olympus Co., Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 6, 2023
- Initiation Date
- October 16, 2023
- Classification Date
- November 29, 2023
- Address
- 3 Chome 1-1, Niiderakita, Aizuwakamatsu, N/A, Japan
Description
GIF-H190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Reason
Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
Code Info
UDI-DI 04953170305290 Serial Numbers 2515895, 2312905, 2628532
Distribution
Domestic distribution to FL, KY, MA, NJ, NY, and PA
Quantity
24 incorrectly repaired units