TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDNS090Z
Enforcement
- Recall Number
- Z-0415-2022
- Event ID
- 89170
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Acclarent, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 29, 2021
- Initiation Date
- November 12, 2021
- Classification Date
- December 22, 2021
- Address
- 31 Technology Dr Suite 200, N/A, Irvine, CA, 92618-2346, United States
Description
TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDNS090Z
Suction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect the accuracy error, this could lead to an intraprocedural delay, a potential for reversible tissue damage or damage to the carotid artery, CSF leak, or damage to optic nerve.
0-Degree Lots: 1903120, 1903250, 1904060, 1904180, 1904240, 1905030, 1905070, 1906260, 1907120, 1907170, 1907220, 1907240, 1907260, 2003040, 2009110, 2011240, 2102080, 2103090, 2104260, 2105040, 2105260, 2106100, 2107070. 70-Degree Lots: 1903137, 1903257, 1904067, 1905017, 1905037, 1905077, 1906067, 1906267, 1907127, 1908137, 1908157, 2003127. 90-Degree: 1903139, 1903259, 1904069, 1905019, 1905139, 1906079, 1908269, 1909279, 1909309, 1910299, 2003029, 2004029
US: AZ, MI, IA, LA, IL, VA, SC, NY, IN, OR, SD, TX, PA, NE, FL, CT, MN, NV, CA, NJ, UT, AL, MA, WI, TN, WV, OH, NC, KY, HI¿, OK, GA, CO, MO, WY, MS, NH, HI, MD, ID