FDA Enforcement Class II Terminated

Touchscreen display kit. Sold under the following product names: TOUCHSCREEN DISPLAY KIT 11.X ENGLISH FOR USA, TOUCHSCREEN DISPLAY KIT 11.X ENGLISH, TOUCHSCREEN DISPLAY KIT 11.X GERMAN, TOUCHSCREEN DISPLAY KIT 11.X DUTCH, TOUCHSCREEN DISPLAY KIT 11.X FRENCH, TOUCHSCREEN DISPLAY KIT 11.X DANISH, TOUCHSCREEN DISPLAY KIT 11.X ESTONIAN, TOUCHSCREEN DISPLAY KIT 11.X ITALIAN, TOUCHSCREEN DISPLAY KIT 11.X NORWEGIAN. Product Usage: The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

Recall: Z-0413-2019 · Reported November 21, 2018

Enforcement

Recall Number
Z-0413-2019
Event ID
81182
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
GE Healthcare, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 21, 2018
Initiation Date
September 27, 2018
Classification Date
November 9, 2018
Termination Date
April 12, 2021
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615, United States

Description

Touchscreen display kit. Sold under the following product names: TOUCHSCREEN DISPLAY KIT 11.X ENGLISH FOR USA, TOUCHSCREEN DISPLAY KIT 11.X ENGLISH, TOUCHSCREEN DISPLAY KIT 11.X GERMAN, TOUCHSCREEN DISPLAY KIT 11.X DUTCH, TOUCHSCREEN DISPLAY KIT 11.X FRENCH, TOUCHSCREEN DISPLAY KIT 11.X DANISH, TOUCHSCREEN DISPLAY KIT 11.X ESTONIAN, TOUCHSCREEN DISPLAY KIT 11.X ITALIAN, TOUCHSCREEN DISPLAY KIT 11.X NORWEGIAN. Product Usage: The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

Reason

Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro Spontaneous Breathing Modes do not provide the feature of transitioning to SIMV PCV backup ventilation mode for a patient who stops spontaneous breathing if the PSVPro ventilation mode is in use prior to the clinician starting the Cycling Procedure.

Code Info

Model Number, UDI Number, (Serial Number): 2097673-001-US (4651698-1); 2097673-001-US (4651698-3); 2097673-001-US (4651698-4); 2097673-001-US (4651726-1); 2097673-001-US (4652206-1); 2097673-001-US (4652206-2); 2097673-001-US (4649618-1); 2097673-001-US (4649618-2); 2097673-001-US (4649618-3); 2097673-001-US (4649618-4); 2097673-001-US (4649618-5); 2097673-001-US (4653227-1); 2097673-001-US (4653227-2); 2097673-001-US (4653227-3); 2097673-001-US (5474296-1); 2097673-001-US (4652587-1); 2097673-001-US (4652587-2); 2097673-001-US (4652587-3); 2097673-001-US (4652587-4); 2097673-001-US (4652587-5); 2097673-001-US (4652587-6); 2097673-001-US (4632861-1); 2097673-001-US (4632861-2); 2097673-001-US (4632861-3); 2097673-001-US (4644953-1); 2097673-001 (5463757-1); 2097673-001 (5467309-16); 2097673-001 (5471699-1); 2097673-001 (5471699-2); 2097673-001 (5471699-3); 2097673-001 (5475515-1); 2097673-001 (5475515-2); 2097673-001 (5475516-1); 2097673-001 (5475516-10); 2097673-001 (5475516-2); 2097673-001 (5475516-3); 2097673-001 (5475516-4); 2097673-001 (5475516-5); 2097673-001 (5475516-6); 2097673-001 (5475516-7); 2097673-001 (5475516-8); 2097673-001 (5475516-9); 2097673-001 (5475528-1); 2097673-001 (5475528-10); 2097673-001 (5475528-11); 2097673-001 (5475528-12); 2097673-001 (5475528-13); 2097673-001 (5475528-14); 2097673-001 (5475528-15); 2097673-001 (5475528-16); 2097673-001 (5475528-17); 2097673-001 (5475528-18); 097673-001 (5475528-19); 2097673-001 (5475528-2); 2097673-001 (5475528-3); 2097673-001 (5475528-4); 2097673-001 (5475528-5); 2097673-001 (5475528-6); 2097673-001 (5475528-7); 2097673-001 (5475528-8); 2097673-001 (5475528-9); 2097673-001 (5481650-1); 2097673-001 (5481650-10); 2097673-001 (5481650-2); 2097673-001 (5481650-3); 097673-001 (5481650-4); 2097673-001 (5481650-5); 2097673-001 (5481650-6); 2097673-001 (5481650-7); 2097673-001 (5481650-8); 2097673-001 (5481650-9); 2097673-001 (5471701-1); 2097673-001 (5471701-2); 2097673-001 (5471702-1); 2097673-001 (5472013-1); 2097673-001 (5472013-10); 097673-001 (5472013-12); 2097673-001 (5472013-14); 2097673-001 (5472013-15); 2097673-001 (5472013-16); 2097673-001 (5472013-17); 2097673-001 (5472013-18); 2097673-001 (5472013-3); 2097673-001 (5472013-4); 2097673-001 (5472013-5); 2097673-001 (5472013-6); 2097673-001 (5472013-7); 2097673-001 (5472013-8); 2097673-001 (5463631-1); 2097673-001 (5463631-2); 2097673-001 (5466892-1); 2097673-001 (5466892-10); 2097673-001 (5466892-2); 2097673-001 (5466892-3); 2097673-001 (5466892-4); 2097673-001 (5466892-5); 2097673-001 (5466892-6); 2097673-001 (5466892-7); 2097673-001 (5466892-8); 2097673-001 (5466892-9); 2097673-DE (5480919-1); 2097673-NL (5473301-1); 2097673-NL (5473301-2); 2097673-NL (5473301-3); 2097673-NL (5473301-4); 2097673-NL (5473301-5); 2097673-NL (5473301-6); 2097673-FR (5481162-1); 2097673-NL (5481369-1); 2097673-NL (5481369-2); 2097673-NL (5481369-3); 2097673-NL (5481369-4); 2097673-NL (5481369-5); 2097673-NL (5481369-6); 2097673-NL (5481369-7); 2097673-NL (5481369-8); 2097673-NL (5481369-9); 2097673-001 (5462738-1); 2097673-001 (5462738-2); 2097673-FR (5475371-1); 2097673-FR (5475371-2); 2097673-FR (5475371-3); 2097673-FR (5475371-4); 2097673-FR (5475371-5); 2097673-FR (5479900-1); 2097673-DA (5453900-1); 2097673-ET (5461749-1); 2097673-FR (5481462-1); 2097673-FR (5481462-2); 2097673-FR (5481462-3); 2097673-FR (5481462-4); 2097673-FR (5481462-5); 2097673-FR (5481462-6); 2097673-FR (5481462-7); 2097673-FR (5481462-8); 2097673-FR (5481462-9); 2097673-FR (5473743-1); 2097673-FR (5473743-2); 2097673-FR (5473743-3); 2097673-FR (5473743-4); 2097673-FR (5473743-5); 2097673-FR (5473744-1); 2097673-FR (5473744-2); 2097673-FR (5473744-3); 2097673-FR (5473744-4); 2097673-FR (5473744-5); 2097673-FR (5473744-6); 2097673-FR (5458277-1); 2097673-FR (5458277-2); 2097673-001 (5455478-1); 2097673-IT (5476554-1); 2097673-IT (5476554-2); 2097673-IT (5476554-3); 2097673-NL (5450447-1); 2097673-NL (5450447-3); 2097673-NL (5450447-5); 2097673-NL (5450447-6); 2097673-NO (5474251-1); 2097673-001 (5476468-70); 2097673-001 (5476468-71); 2097673-001 (5476468-72); 2097673-001 (5476468-73); 2097673-001 (5476468-74); 2097673-001 (5476468-75); 2097673-001 (5476468-76); 2097673-001 (5476468-77); 2097673-001 (5476468-78); 2097673-001 (5476468-79); 2097673-001 (5476468-80); 2097673-001 (5476468-81); 2097673-001 (5476468-82); 2097673-001 (5476468-83); 2097673-001 (5476468-84); 2097673-001 (5476468-85); 2097673-001 (5476468-86); 2097673-001-US (4673852-1); 2097673-001-US (4673852-2); 2097673-001-US (4673852-3); Additional Serial Numbers - Model Number, UDI Number, (Serial Number): 2097673-001-US, Not Applicable, (ANAM00306); 2097673-001-US, Not Applicable, (ANAP00354); 2097673-001-US, Not Applicable, (APWU00297); 2097673-001-US, Not Applicable, (ANAP00715); 2097673-001-US, Not Applicable, (ANAR00602); 2097673-001-US, Not Applicable, (ANAR00603); 2097673-001-US, Not Applicable, (ANAN00169); 2097673-001-US, Not Applicable, (ANAP00574); 2097673-001-US, Not Applicable, (ANAP00575); 2097673-001-US, Not Applicable, (ANAP00576); 2097673-001-US, Not Applicable, (ANAP00577); 2097673-001-US, Not Applicable, (ANAQ01175); 2097673-001-US, Not Applicable, (ANAP00172); 2097673-001-US, Not Applicable, (ANAP00861); 2097673-001-US, Not Applicable, (ANAP00870); 2097673-001-US, Not Applicable, (ANAP00872); 2097673-001-US, Not Applicable, (ANAS00398 ); 2097673-001-US, Not Applicable, (ANAL00105); 2097673-001-US, Not Applicable, (ANAL00106); 2097673-001-US, Not Applicable, (ANAL00107); 2097673-001-US, Not Applicable, (ANAL00108); 2097673-001-US, Not Applicable, (ANAL00109); 2097673-001-US, Not Applicable, (ANAL00116); 2097673-001-US, Not Applicable, (ANAM00439); 2097673-001-US, Not Applicable, (ANAM00440); 2097673-001-US, Not Applicable, (ANAM00441); 2097673-001-US, Not Applicable, (ANAP01088); 2097673-001-US, Not Applicable, (ANAP01090); 2097673-001-US, Not Applicable, (ANAP01091); 2097673-001, Not Applicable, (ANAN00604); 2097673-001, Not Applicable, (ANAM01165); 2097673-001, Not Applicable, (ANAM01158); 2097673-001, Not Applicable, (ANAM00815); 2097673-001, Not Applicable, (ANAM00816); 2097673-001, Not Applicable, (ANAM00817); 2097673-001, Not Applicable, (ANAM00818); 2097673-001, Not Applicable, (ANAM00819); 2097673-001, Not Applicable, (ANAM00820); 2097673-001, Not Applicable, (ANAM00821); 2097673-001, Not Applicable, (ANAM00822); 2097673-001, Not Applicable, (ANAM00823); 2097673-001, Not Applicable, (ANAM00824); 2097673-DE, Not Applicable, (ANAP01156); 2097673-001, Not Applicable, (ANAR00318); 2097673-001, Not Applicable, (ANAR00319); 2097673-FR, Not Applicable, (ANAR05232); 2097673-FR, Not Applicable, (ANAR05233); 2097673-FR, Not Applicable, (ANAR05234); 2097673-FR, Not Applicable, (ANAR05235); 2097673-FR, Not Applicable, (ANAR05236); 2097673-FR, Not Applicable, (ANAP01419); 1011-9000-000, Not Applicable, (APWT00661); 1011-9000-000, Not Applicable, (APWT00662); 1011-9000-000, Not Applicable, (APWT00664); 1011-9000-000, Not Applicable, (ANAN00664); 1011-9000-000, Not Applicable, (ANAN00666); 1011-9000-000, Not Applicable, (ANAN00667); 1011-9000-000, Not Applicable, (ANAN00669); 1011-9000-000, Not Applicable, (ANAN00670); 1011-9000-000, Not Applicable, (ANAN00671); 1011-9000-000, Not Applicable, (ANAN00890); 1011-9000-000, Not Applicable, (ANAN00891); 1011-9000-000, Not Applicable, (ANAR00396); 1011-9000-000, Not Applicable, (APWT02012); 1011-9000-000, Not Applicable, (APWT02013); 1011-9000-000, Not Applicable, (ANAN01101); 1011-9000-000, Not Applicable, (ANAP00605); 2097673-FR, Not Applicable, (ANAQ00803); 2097673-FR, Not Applicable, (ANAQ00806); 2097673-FR, Not Applicable, (ANAQ00948); 2097673-FR, Not Applicable, (ANAR05145); 2097673-FR, Not Applicable, (ANAR05150); 2097673-FR, Not Applicable, (ANAR05513); 2097673-FR, Not Applicable, (ANAR05517); 2097673-FR, Not Applicable, (ANAS00665); 2097673-FR, Not Applicable, (ANAS00666); 2097673-FR, Not Applicable, (ANAS00668); 2097673-FR, Not Applicable, (ANAS00971); 2097673-NO, Not Applicable, (ANAQ01098);

Distribution

Worldwide distribution - US nationwide in the states of CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, ,MA, MD, MI, MN, MO, MT, NC, NE, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV. Countries of Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (Republic of), Kuwait, Macedonia, Mexico, Netherlands, New Zealand, Nigeria, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.

Quantity

4,207 units total