Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: Ultrasound device for intraoperative use Component: No
Enforcement
- Recall Number
- Z-0410-2026
- Event ID
- 97707
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medistim Asa
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 5, 2025
- Initiation Date
- September 26, 2025
- Classification Date
- October 29, 2025
- Address
- Okernveien 94, N/A, Oslo, N/A, N/A, Norway
Description
Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011, MQV12011 Software Version: Not applicable Product Description: Ultrasound device for intraoperative use Component: No
Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.
Lot Code: Serial number; Model number 5042; MQC04001 5043; MQC04001 5044; MQC14001 5061; MQU14011 5062; MQU14011 5064; MQU14011 5065; MQU14011 5066; MQU14011 5067; MQC04001 5037; MQU14011 5073; MQV12011 5038; MQU14011 5045; MQU14011 5063; MQU14011
Domestic distribution to CO, MI, MN, TN, TX. International distribution to Australia, Austria, Germany, Finland, India, Japan, Russia, China, Saudi Arabia.
14