FDA Enforcement Class II Terminated

Brilliance 64 Computed Tomography X-Ray System, Model number 728231, running version 2.6.0. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Recall: Z-0409-2016 · Reported December 16, 2015

Enforcement

Recall Number
Z-0409-2016
Event ID
69055
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 16, 2015
Initiation Date
March 3, 2011
Classification Date
December 10, 2015
Termination Date
December 15, 2015
Address
595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States

Description

Brilliance 64 Computed Tomography X-Ray System, Model number 728231, running version 2.6.0. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Reason

Multiple software defects which could impact the performance or reliability of the system. These impacts include, but are not limited to: error messages which prevent operation, incorrect scan region, couch movement failure, inaccurate CTDI reporting, image artifacts, lost images. There is a risk of misdiagnosis.

Code Info

Serial numbers: 4001, 4020, 4021, 9030, 9033, 9040, 9041, 9048, 9049, 9062, 9074, 9083, 9093, 9105, 9106, 9107, 9108, 9112, 9123, 9126, 9127, 9131, 9135, 9137, 9139, 9140, 9142, 9148, 9151, 9159, 9160, 9162, 9163, 9166, 9171, 9173, 9175, 9177, 9180, 9196, 9197, 9201, 9202, 9205, 9206, 9208, 9214, 9215, 9216, 9217, 9218, 9235, 9501, 9502, 9503, 9504, 9505, 9506, 9508, 9509, 9510, 9511, 9512, 9513, 9514, 9515, 9516, 9517, 9518, 9519, 9520, 9521, 9522, 9523, 9524, 9525, 9526, 9527, 9528, 9529, 9530, 9531, 9532, 9533, 9535, 9536, 9537, 9538, 9539, 9540, 9541, 9542,9543, 9544, 9545, 9546, 9547, 9548, 9549, 9550, 9551, 9552, 9553, 9554, 9555, 9556, 9557, 9558, 9559, 9560, 9561, 9562, 9563, 9564, 9565, 9566, 9567, 9568, 9569, 9570, 9571, 9572, 9573, 9574, 9575, 9576, 9577, 9578, 9579, 9580, 9581, 9582, 9583, 9584, 9585, 9586, 9587, 9588, 9589, 9590, 9591, 9592, 9593, 9594, 9595, 9596, 9597, 9598, 9599, 9600, 9601, 9602, 9603, 9604, 9605, 9606, 9607, 9608, 9610, 9611, 9612, 9613, 9614, 9615, 9616, 9617, 9618, 9619, 9620, 9621, 9622, 9623, 9624, 9625, 9626, 9627, 9628, 9629, 9630, 9631, 9632, 9633, 9634, 9635, 9636, 9637, 9638, 9639, 9640, 9641, 9642, 9643, 9644, 9645, 9646, 9647, 9648, 9649, 9650, 9651, 9652, 9653, 9654, 9655, 9656, 9658, 9659, 9660, 9661, 9662, 9663, 9664, 9665, 9666, 9667, 9668, 9669, 9670, 9671, 9672, 9673, 9674, 9675, 9676, 9677, 9678, 9679, 9680, 9681, 9682, 9683, 9684, 9685, 9686, 9687, 9688, 9689, 9690, 9691, 9692, 9693, 9694, 9695, 9696, 9697, 9698, 9699, 9700, 9701, 9702, 9703, 9704, 9705, 9706, 9707, 9708, 9709, 9710, 9711, 9712, 9713, 9714, 9715, 9716, 9717, 9718, 9719, 9720, 9721, 9722, 9723, 9724, 9725, 9726, 9727, 9728, 9729, 9730, 9731, 9732, 9733, 9734, 9735, 9736, 9737, 9738, 9739, 9740, 9741, 9742, 9743, 9744, 9745, 9746, 9747, 9748, 9749, 9750, 9751, 9752, 9753, 9754, 9755, 9756, 9757, 9758, 9759, 9760, 9761, 9762, 9763, 9764, 9765, 9766, 9767, 9768, 9769, 9770, 9771, 9772, 9773, 9774, 9775, 9776, 9777, 9778, 9779, 9780, 9781, 9782, 9783, 9784, 9785, 9786, 9787, 9788, 9789, 9790, 9791, 9792, 9793, 9794, 9795, 9796, 9797, 9798, 9799, 9800, 9801, 9802, 9803, 9804, 9805, 9806, 9807, 9808, 9809, 9810, 9811, 9812, 9813, 9814, 9815, 9816, 9817, 9818, 9819, 9820, 9821, 9822, 9823, 9824, 9825, 9826, 9827, 9828, 9829, 9830, 9831, 9832, 9833, 9834, 9835, 9836, 9837, 9838, 9839, 9840, 9841, 9842, 9843, 9844, 9845, 9846, 9847, 9848, 9850, 9851, 9852, 9853, 9854, 9855, 9856, 9857, 9858, 9859, 9860, 9861, 9862, 9864, 9865, 9866, 9867, 9868, 9869, 9870, 9871, 9872, 9873, 9874, 9875, 9876, 9877, 9878, 9879, 9880, 9881, 9882, 9883, 9884, 9885, 9886, 9887, 9888, 9889, 9890, 9891, 9892, 9893, 9894, 9895, 9897, 9898, 9900, 9901, 9902, 9903, 9904, 9905, 9906, 9907, 9908, 9909, 9910, 9911, 9912, 9913, 9914, 9915, 9916, 9917, 9918, 9919, 9920, 9921, 9922, 9923, 9924, 9925, 9926, 9927, 9929, 9930, 9931, 9932, 9933, 9934, 9935, 9936, 9937, 9938, 9940, 9941, 9942, 9946, 9947, 9948, 9950, 9951, 9953, 9954, 9955, 9956, 9958, 9960, 9961, 9962, 9963, 9965, 9968, 9969, 9970, 9971, 9972, 9974, 9975, 9976, 9977, 9978, 9979, 9980, 9981, 9983, 9984, 9985, 9986, 9987, 9989, 9993, 9994, 9995, 9996, 9997, 9999, 10000, 10001, 10002, 10003, 10004, 10007, 10009, 10011, 10012, 10013, 10014, 10015, 10016, 10018, 10021, 10023, 10027, 10028, 10032, 10033, 10034, 10035, 10036, 10037, 10039, 10040, 10041, 10042, 10044, 10045, 10046, 10047, 10048, 10050, 10051, 10052, 10053, 10054, 10057, 10058, 10065, 10066, 10067, 10070, 10072, 10076, 10077, 10078, 10079, 10080, 10081, 10084, 10085, 10086, 10087, 10090, 10091, 10092, 10094, 10095, 10096, 10098, 10099, 10100, 10101, 10102, 10103, 10105, 10106, 10107, 10108, 10110, 10111, 10112, 10113, 10114, 10115, 10116, 10117, 10118, 10119, 10120, 10121, 10122, 10123, 10124, 10125, 10126, 10127, 10128, 10129, 10130, 10131, 10132, 10133, 10134, 10135, 10136, 10137, 10138, 10139, 10140, 10141, 10142, 10143, 10144, 10145, 10146, 10147, 10148, 10149, 10150, 10151, 10152, 10153, 10154, 10155, 10156, 10157, 10158, 10159, 10160, 10161, 10162,10163, 10164, 10165, 10166, 10167, 10168, 10169, 10170, 10171, 10172, 10173, 10174, 10175, 10176, 10177, 10178, 10179, 10180, 10181, 10182, 10183, 10184, 10185, 10186, 10187, 10188, 10189, 10190, 10191, 10192, 10193, 10195, 10196, 10197, 10198, 10200, 10203, 10204, 10206, 10207, 10208, 10209, 10210, 10211, 10213, 10216, 10218, 10221, 10222, 10223, 10224, 10225, 10226, 10229, 10230, 10232, 10233, 10234, 10237, 10238, 10239, 10240, 10241, 10242, 10243, 10244, 10245, 10246, 10247, 10248, 10250, 10252, 10255, 10256, 10257, 10258, 10259, 10260, 10262, 10263, 10264, 10267, 10268, 10270, 10271, 10272, 10273, 10274, 10275, 10276, 10277, 10278, 10279, 10280, 10281, 10282, 10283, 10284, 10285, 10286, 10287, 10288, 10291, 10293, 10294, 10295, 10296, 10297, 10298, 10299, 10300, 10301, 10302, 10303, 10304, 10305, 10307, 10308, 10309, 10311, 10312, 10313, 10315, 10316, 10318, 10319, 10320, 10326, 10330, 10332, 10333, 10334, 10335, 10336, 10337, 10338, 10339, 10341, 10342, 10343, 10344, 10345, 10346, 10347, 10348, 10349, 10350, 10351, 10352, 10353, 10354, 10355, 10356, 10357, 10358, 10359, 10360, 10361, 10362, 10363, 10364, 10365, 10366, 10367, 10368, 10369, 10370, 10371, 10372, 10373, 10374, 10376, 10377, 10380, 10381, 10382, 10383, 10384, 10385, 10386, 10387, 10388, 10389, 10390, 10391, 10392, 10393, 10394, 10395, 10396, 10397, 10398, 10399, 10400, 10401, 10402, 10403, 10404, 10406, 10407, 10408, 10409, 10410, 10411, 10412, 10413, 10414, 10417, 10418, 10420, 10421, 10422, 10423, 10424, 10425, 10426, 10427, 10428, 10429, 10430, 10431, 10432, 10433, 10434, 10436, 10437, 10438, 10439, 29002, 29003, 29005, 29032, 29035, 29045, 29056, 29066, 29092, 29107, 64015, 64173, 65006, 85210, 90029, 90030, 90031, 90033, 90036, 90037, 90042, 90045, 90050, 90054, 90055, 90057, 90060, 90062, 90063, 90066, 90070, 90072, 90073, 90074, 90077, 90078, 90080, 90081, 90082, 90084, 90085, 90086, 90087, 90089, 90091, 90092, 90094, 90096, 90097, 90098, 90099, 90100, 90105, 90107, 90117, 90118, 90119, 90120, 90123, 90124, 90129, 90131, 90132, 90135, 90136, 90137, 90139, 90140, 90141, 90143, 90147, 90151, 90152, 90153, 90154, 90156, 90159, 90160, 90161, 90162, 90163, 90167, 90168, 90170, 90175, 90177, 90182, 90205, 90218, 91003, 92029, 94134, 95003, 95008, 95009, 95011, 95012, 95013, 95014, 95015, 95016, 95017, 95018, 95019, 95020, 95021, 95022, 95023, 95024, 95025, 95026, 95029, 95030, 95031, 95032, 95033, 95034, 95035, 95036, 95037, 95038, 95039, 95040, 95041, 95042, 95043, 95044, 95045, 95046, 95047, 95048, 95049, 95050, 95051, 95052, 95053, 95055, 95056, 95057, 95058, 95059, 95060, 95061, 95062, 95063, 95064, 95065, 95066, 95067, 95068, 95069, 95071, 95072, 95073, 95074, 95075, 95076, 95077, 95078, 95079, 95080, 95081, 95082, 95083, 95084, 95085, 95086, 95087, 95088, 95089, 95090, 95091, 95092, 95093, 95094, 95095, 95096, 95097, 95098, 95099, 95100, 95101, 95102, 95103, 95104, 95105, 95106, 95107, 95108, 95109, 95110, 95111, 95112, 95113, 95114, 95116, 95117, 95118, 95119, 95120, 95121, 95122, 95123, 95124, 95125, 95126, 95127, 95128, 95129, 95130, 95131, 95132, 95133, 95134, 95135, 95136, 95137, 95138, 95139, 95140, 95141, 95142, 95143, 95144, 95145, 95146, 95147, 95148, 95149, 95150, 95151, 95152, 95153, 95154, 95155, 95156, 95157, 95158, 95159, 95160, 95161, 95163, 95164, 95165, 95166, 95167, 95168, 95169, 95170, 95171, 95172, 95173, 95174, 95175, 95176, 95177, 95178, 95179, 95180, 95181, 95182, 95183, 95184, 95185, 95186, 95187, 95188, 95189, 95190, 95191, 95192, 95193, 95194, 95195, 95196, 95197, 95198, 95199, 95200, 95201, 95202, 95203, 95204, 95205, 95206, 95207, 95208, 95209, 95210, 95211, 95212, 95213, 95214, 95215, 95216, 95217, 95218, 95219, 95220, 95221, 95222, 95223, 95224, 95225, 95226, 95227, 95228, 95229, 95231, 95232, 95233, 95234, 95235, 95236, 95237, 95238, 95239, 95240, 95241, 95242, 95243, 95244, 95245, 95246, 95247, 95249, 95250, 95251, 95252, 95253, 95254, 95255, 95256, 95257, 95258, 95259, 95260, 95261, 95262, 95263, 95264, 95265, 95266, 95267, 95268, 95269, 95270, 95271, 95272, 95273, 95274, 95275, 95276, 95277, 95278, 95279, 95280, 95281, 95282, 95283, 95284, 95285, 95286, 95287, 95288, 95289, 95290, 95291, 95292, 95293, 95294, 95295, 95296, 95297, 95298, 95299, 95300, 95301, 95302, 95303, 95304, 95305, 95307, 95308, 95309, 95310, 95311, 95312, 95313, 95314, 95315, 95316, 95317, 95318, 95319, 95320, 95321, 95322, 95323, 95324, 95325, 95326, 95327, 95328, 95329, 95330, 95331, 95332, 95333, 95334, 95335, 95336, 95337, 95338, 95339, 95340, 95341, 95342, 95343, 95344, 95345, 95346, 95347, 95348, 95349, 95350, 95351, 95352, 95353, 95354, 95355, 95356, 95357, 95358, 95359, 95360, 95361, 95362, 95363, 95364, 95365, 95366, 95367, 95368, 95369, 95370, 95371, 95373, 95374, 95375, 95376, 95377, 95378, 95379, 95380, 95381, 95382, 95383, 95384, 95385, 95386, 95387, 95388, 95389, 95390, 95391, 95393, 95394, 95395, 95397, 95398, 95399, 95400, 95401, 95402, 95403, 95404, 95405, 95406, 95407, 95408, 95409, 95410, 95411, 95412, 95413, 95414, 95415, 95416, 95417, 95418, 95419, 95420, 95421, 95422, 95423, 95424, 95425, 95426, 95427, 95428, 95429, 95430, 95431, 95432, 95433, 95434, 95435, 95436, 95437, 95438, 95439, 95440, 95441, 95443, 95444, 95445, 95446, 95447, 95448, 95449, 95450, 95451, 95452, 95453, 95454, 95455, 95456, 95457, 95458, 95459, 95460, 95461, 95462, 95463, 95464, 95465, 95466, 95467, 95468, 95469, 95470, 95471, 95472, 95473, 95474, 95475, 95476, 95477, 95478, 95479, 95480, 95481, 95482, 95483, 95484, 95485, 95486, 95487, 95488, 95489, 95490, 95491, 95492, 95493, 95494, 95495, 95496, 95497, 95498, 95499, 95500, 95501, 95502, 95503, 95504, 95505, 95506, 95507, 95508, 95509, 95510, 95511, 95512, 95513, 95514, 95515, 95516, 95517, 95518, 95519, 95520, 95521, 95522, 95523, 95524, 95525, 95526, 95527, 95528, 95529, 95530, 95531, 95532, 95533, 95534, 95535, 95536, 95537, 95538, 95539, 95540, 95541, 95542, 95543, 95544, 95545, 95546, 95547, 95548, 95549, 95550, 95551, 95552, 95553, 95554, 95555, 95556, 95557, 95558, 95559, 95560, 95561, 95562, 95563, 95564, 95565, 95566, 95568, , 5569, 95570, 95572, 95573, 95574, 95575, 95577, 95578, 95579, 95580, 95581, 95582, 95583, 95584, 95585, 95586, 95587, 95588, 95589, 95590, 95591, 95592, 95593, 95594, 95595, 95596, 95597, 95598, 95599, 95600, 95601, 95602, 95603, 95604, 95605, 95606, 95607, 95608, 95609, 95610, 95611, 95612, 95613, 95614, 95615, 95616, 95617, 95618, 95619, 95620, 95621, 95622, 95623, 95624, 95625, 95626, 95627, 95628, 95629, 95630, 95631, 95632, 95633, 95634, 95635, 95636, 95637, 95638, 95639, 95640, 95641, 95642, 95643, 95644, 95645, 95646, 95647, 95648, 95649, 95650, 95651, 95652, 95653, 95654, 95655, 95656, 95657, 95658, 95659, 95660, 95661, 95662, 95663, 95664, 95665, 95666, 95667, 95668, 95669, 95670, 95671, 95672, 95673, 95674, 95675, 95676, 95677, 95678, 95679, 95680, 95681, 95682, 95683, 95684, 95685, 95686, 95687, 95688, 95689, 95690, 95691, 95692, 95693, 95694, 95695, 95696, 95697 and 410249.

Distribution

Worldwide Distribution. US nationwide, including Puerto Rico, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Equatorial Guinea, Finland, France, Germany, Greece, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia & Montenegro, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam & Yemen.

Quantity

1,765 units