FDA Enforcement Class II Terminated

PURE OXYGEN CONCENTRATOR, CHAD Drive, Models CH5000 and CH5000S

Recall: Z-0407-2018 · Reported January 31, 2018

Enforcement

Recall Number
Z-0407-2018
Event ID
78883
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Inovo, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 31, 2018
Initiation Date
October 9, 2017
Classification Date
January 24, 2018
Termination Date
September 15, 2023
Address
401 Leonard Blvd N, N/A, Lehigh Acres, FL, 33971-6302, United States

Description

PURE OXYGEN CONCENTRATOR, CHAD Drive, Models CH5000 and CH5000S

Reason

Post-market surveillance for Pure Stationary Oxygen Concentrators model CHSOOO and CHSOOOS demonstrate an aggregate complaint/service rate greater than 35% for low oxygen purity (output oxygen below the published specification).

Code Info

UDI #(s): 00814470020013 (model CHSOOO) and 00814470020006 (model CHSOOOS): All serial numbers

Distribution

worldwide

Quantity

13,140 units