FDA Enforcement
Class II
Ongoing
with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system.
Recall: Z-0405-2024
·
Reported December 6, 2023
Enforcement
- Recall Number
- Z-0405-2024
- Event ID
- 93331
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Breas Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- December 6, 2023
- Initiation Date
- October 13, 2023
- Classification Date
- November 27, 2023
- Address
- 16 Esquire Rd, North Billerica, MA, 01862-2527, United States
Description
with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system.
Reason
Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation
Code Info
UDI-DI: 05030267120505 NIV Angled Mouthpiece with Notch 22M/15F; 07321820060931 MPV Kit 006093; 07321820066537 MPV Kit 006653 MPV mouthpiece Lot Numbers: 32304444, 32305269 and 32306017
Distribution
Nationwide Foreign: Canada
Quantity
215 units