FDA Enforcement Class II Terminated

Freedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System is a Class I Medical Device. The device listing number this product is E115952. The intended use of this device is to illuminate oral structures and operating areas.

Recall: Z-0402-2013 · Reported November 28, 2012

Enforcement

Recall Number
Z-0402-2013
Event ID
63512
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Kerr Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 28, 2012
Initiation Date
November 17, 2011
Classification Date
November 20, 2012
Termination Date
November 20, 2012
Address
3225 Deming Way, N/A, Middleton, WI, 53562, United States

Description

Freedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System is a Class I Medical Device. The device listing number this product is E115952. The intended use of this device is to illuminate oral structures and operating areas.

Reason

The firm recalled the Freedom Cordless LED Headlight because some units were assembled with an incorrect component in the control board electronics that may cause some Freedom Cordless LED Headlight units to become warm to the touch; however, no skin damage will occur.

Code Info

Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System has no expiration date. The affected units were produced and distributed since February 2011.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, and Europe.

Quantity

1330 units