CP5 Centrifugal Pump, made up of the drive unit, the emergency system and the pump control panel. It used with an Electrical Remote-Controlled (ERC) tubing clamp and a bubble sensor.
Enforcement
- Recall Number
- Z-0401-2018
- Event ID
- 78594
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Sorin Group Deutschland GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 31, 2018
- Initiation Date
- October 30, 2017
- Classification Date
- January 19, 2018
- Termination Date
- January 26, 2024
- Address
- Lindberghstr. 25, N/A, Munich, N/A, N/A, Germany
Description
CP5 Centrifugal Pump, made up of the drive unit, the emergency system and the pump control panel. It used with an Electrical Remote-Controlled (ERC) tubing clamp and a bubble sensor.
Possibility that the automatic line closure by the Electronic Remote-Controlled (ERC) clamp after detection of an air bubble could take longer than is intended by design in the CP5 system.
Model No. 60-00-60. Drive Unit: Ref #60-01-04, UDI 04033817900948. Emergency System: Ref #60-01-35, UDI 04033817900955. Pump Control Panel: Ref #60-02-60, UDI 04033817901006.
Distributed domestically to Distributed internationally to Algeria, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Jordan, Kuwait, Libya, Lithuania, Luxembourg, Malaysia, Morocco, Norway, Palestinian Territory, Panama, Peru, Poland, Qatar, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, The Netherlands, Tunisia, United Kingdom, Vietnam
1,318 units