SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825
Enforcement
- Recall Number
- Z-0399-2022
- Event ID
- 89035
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 29, 2021
- Initiation Date
- November 10, 2021
- Classification Date
- December 20, 2021
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825
Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.
Lot: M106450, UDI: 15099590575113
US Nationwide distribution in the states of Louisiana, Ohio, North Carolina, California, South Carolina, Oklahoma, Pennsylvania, New York, Florida, New Jersey, Wisconsin, Alabama, Texas, Georgia, Maine, North Dakota, Virginia, Kansas, Washington, New Mexico, Iowa ,Nebraska, Idaho, Michigan, Maryland, Arizona, Hawaii, Missouri, Oregon, Mississippi, Rhode Island, Alaska, Tennessee, Illinois, Delaware, Kentucky, Arkansas, West Virginia, Massachusetts. OUS: India, South Africa, Saudi Arabia, Canada, Lebanon, Singapore, Philippines, Thailand