StageOne Hip, Models 431207 & 431209Usage: Product Usage: The single-use cement spacer molds are sterile disposables made of medical grade silicone with a 316L stainless steel reinforcement stem. They are designed to be filled with polymethylmethacrylate/gentamicin bone cement, or equivalent, by injecting with a dispenser/gun into the mold. After the cement cures, the temporary spacers are to be removed from the molds with the reinforcement remaining as the core of the spacer, and placed into the joint space. The spacers remain in place (180 days or less) until the second stage of the two-stage procedure is performed to implant a conventional hip joint prosthesis.
Enforcement
- Recall Number
- Z-0399-2019
- Event ID
- 81185
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 14, 2018
- Initiation Date
- September 21, 2018
- Classification Date
- November 7, 2018
- Termination Date
- April 14, 2020
- Address
- 56 E Bell Dr, Warsaw, IN, 46582-6989, United States
Description
StageOne Hip, Models 431207 & 431209Usage: Product Usage: The single-use cement spacer molds are sterile disposables made of medical grade silicone with a 316L stainless steel reinforcement stem. They are designed to be filled with polymethylmethacrylate/gentamicin bone cement, or equivalent, by injecting with a dispenser/gun into the mold. After the cement cures, the temporary spacers are to be removed from the molds with the reinforcement remaining as the core of the spacer, and placed into the joint space. The spacers remain in place (180 days or less) until the second stage of the two-stage procedure is performed to implant a conventional hip joint prosthesis.
There is a potential for comingling of the 43MM and 51 MM molds
Item Number 431207 Lot # 705550, UDI - (01)00880304447165(17)270927(10)705550 Item Number 431209 Lot # 705570, UDI - (01)00880304447141(17)270916(10)705570 Lot # 705580, UDI - (01)00880304447141(17)270916(10)705580
US Nationwide Distribution in the states of: AZ, FL, MI, OH, VA & WI
72 units